| Feature |
A Medical Electronics Manufacturing Fall 1997 Feature
Using high-frequency design methods along with PCB layout and CM filtering results in medical devices that can withstand immunity threats.
With the advent of recent national and international immunity standards, manufacturers of medical electronic equipment have become increasingly concerned about how to design sensitive devices to withstand various transient and steady-state radio-frequency (RF) immunity threats. Emission issues are also a major concern, especially with the use of higher-speed clock circuits.
Electrostatic discharge test (IEC 801-2: 1991) being performed according to EN 60601-1-2.
While many feel that the measures used for emission control will also solve all of the immunity hardening requirements, this is often not the case. Sensitive analog circuits coupled with restrictive leakage current and special external surface treatments for the case have resulted in numerous analog instrument amplifiers and comparator circuits being shut down in the presence of radiated field levels of less than 1 V/m.
Patient interface leakage requirements also make the typical electromagnetic compatibility (EMC) design techniques of signal cable decoupling and cable shielding less effective. This problem has created new solutions in terms of developing effective hardening methods. The key often lies in the printed circuit board (PCB) layout and internal interconnect choices. Thinking in terms of transmission line and high-frequency impedance control methods, even when designing low-frequency analog circuits, can result in effective EMC solutions.
The Threat
The requirements most designers have to deal with today center on FDA's EMC Reviewer Guidance for Premarket Notification Submissions and the European CE requirements that will center on EN 60601-1-2, which becomes mandatory in 1998. Until then, a dual path exists of following the generic EMC/EMI requirements of EN 60601-1-2, or EN 50081 and EN 50082. The basic requirements are identified in the sample test plan shown in Table I. Additional requirementssuch as power-line magnetic fields immunity, power flicker, and power-line surges and sagsare under consideration and may become mandatory in the future.
| Test in Accordance to | Level | Criteria |
|---|---|---|
| IEC 801-3 or ENV 50140 | 3 V/m, 26 to 500 MHz, or 80 MHz to 1.0 GHz modulation | EN 60601-1-2 (Equipment will operate as intended after the test. No degradation of performance is allowed below the specified level provided by the manufacturer.) |
| EN V50141 | 3 V 150 kHz to 80 MHz at 1 kHz modulation onto ac power and I/O cables | EN 60601-1-2 (Equipment will operate as intended after the test. No degradation of performance is allowed below the specified level provided by the manufacturer.) |
| IEC 801-2 (1991) | 8 kV air discharge; 3 kV contact and coupling planes | EN 60601-1-2 (Equipment will operate as intended after the test. No degradation of performance is allowed below the specified level provided by the manufacturer.) |
| IEC 801-4 (1 ed. 1998) | 1 kV at the mains plug; 5/50 pulse at 5 kHz rep rate | EN 60601-1-2 (Equipment will operate as intended after the test. No degradation of performance is allowed below the specified level provided by the manufacturer.) |
| IEC 801-5 | 1 kV differential mode 2 kV common mode | EN 60601-1-2 (Equipment will operate as intended after the test. No degradation of performance is allowed below the specified level provided by the manufacturer.) |
| EN 60555 part 2 | Per customer specification in technical sheet | EN 60555-2 (Equipment will operate as intended after the test. No degradation of performance is allowed below the specified level provided by the manufacturer.) |
| EN 55011 (1991) | Class B | Device to be at least 2 dB under Class B limits. |
Table I. Typical medical test plan, using EN 60601-1-2 (1993) and/or EN 50082-1 (1997) for immunity, and EN 55011 and EN 60555 part 2 (harmonics) for emissions.
FDA premarket notification requirements are far more comprehensive and represent special test methods that result in even higher immunity requirements. The agency borrowed a number of susceptibility requirements from the military and then added further test method restrictions to make them even more severe. When a manufacturer needs to file for 510(k) premarket notification covered by these EMC guidelines, compliance will need to be shown.
Estimating the Amount of Protection
Knowing the amount of protection given circuits will need to meet these requirements in terms of case shielding, circuit filtering, cable protection, and other methods is key. A number of software programs and bookssuch as EMCad (800/500-4EMC) and Albert Smith's book, Electromagnetic Coupling into Transmission Lines by Wyle Inter-sciencecan help with this assessment.
The following steps provide an example of how a quick approximation can be calculated.
This is only an approximation, and more rigorous analysis is recommended. The induced voltage range for this example varied from a factor of 1/150 to 1/80, depending on load impedance, and represents an estimate of the amount of induced voltage across the load impedance. Comparing the analog threshold value to this value will give an estimate of the amount of protection necessary. The needed protection can consist of: cable shield protection over the frequency range of the threat; cable shield termination protection; I/O L-C low-pass lead filtering at port-of-case entry; I/O CM choke filter (external or internal); differential line drivers and receivers; and optic isolators.
Figure 1. Radiated susceptibility of cables (plane wave). Amplitude in dBµV.
The same analysis performed using EMCad yields a plotted result as shown in Figure 1. This analysis shows that the induced voltage varies with frequency and, for 100-
load and source impedance, will induce 118 dBµV, or 0.79 V, at the half-wavelength frequency associated with a 1-m external cable length (assumes no shielding). Comparing this level with the threshold value of 5 mV yields a 44-dB protection need.
Circuit Response
The amplifier response to a 26-MHz threat at first seems unreasonable, considering a typical instrumentation bandwidth capability. Internal IC protection diodes have the ability to dc-rectify the out-of-band RF induced signal and cause the amplifier to be dc-biased into a saturated state. Another possibility is for the nonlinear action of the diodes to block the RF but allow the modulation on the carrier (required to be 1 kHz) to be induced along with the true intended signal.
Possible solutions to out-of-band problems and their typical performance ranges are shown in Table II.
| Protection Method | Minimum Performance (dB) | Maximum Performance (dB) | Comments |
|---|---|---|---|
| Braided shield over cable | 20 | 70 | Assumes shield grounded at both ends* |
| Foil and braid over cable | 30 | 80 | Assumes shield grounded at both ends* |
| Filtering at case entry point | 0 | 100 | Maximum depends on filter type |
| External CM cable choke | 0 | 20 | Single turn on Fair rite #43 material |
| Differential driver/receiver pair | 10 | 80 | Poor at stopping >30-MHz threats |
| Optic isolators | 10 | 80 | Poor at stopping >30-MHz threats |
| *All external cables routed between two shielded cases must have their shield grounded at the point of entry to each shielded case to maintain the shielding integrity of the case. Since dc grounding can result in induced shield current that can couple to the circuit, one end may be ac terminated to case via capacitors. A minimum of two capacitors will need to be inserted 180° apart. | |||
Table II. Possible solutions to out-of-band problems and typical performance ranges.
Hardening Analog Circuits
Methods for hardening analog circuits are as follows:
Controlling Filter Capacitance Values
If signal leads are going to employ line-to-chassis capacitive filtering, then the amount of capacitance will be limited by the allowed leakage current. LC filters where the CM inductance is large will allow the pin-to-case capacitance to be small and stil l achieve the low pass filter insertion loss values needed to protect most circuits. In most cases, the line-to-chassis impedance of bridge and amplifier circuits is very high. Therefore, even a small amount of capacitance will be effective in protecting the circuit.
PCB Designs
The PCB design should provide for a full ground and power plane just like digital boards, and the mother or backplane board should ground all return planes together. In turn, these planes should be capacitively grounded to chassis with capacitive values selected according to the amount of case leakage permitted. The key to grounding is that the ground path must have minimum inductance. The best path is a short (< 0.5 in. long) conductivity-finished metallic standoff located at each mounting point. By chassis grounding the plane area just under the mounting studs and providing a small isolation to the full ground plane, a capacitor, a dc shunt, or nothing can be connected across the signal to chassis ground planes, giving the designer maximum flexibility in grounding.
If dc grounding is employed, the power supply should float power return reference from its chassis, providing one chassis connection reference for the system, namely at the mother or backplane board.
Transient Threats
The electrostatic discharge (ESD) and electrical fast transients (EFTs) are generally more of a concern to digital circuits than to analog circuits. The ESD event generates significant amounts of RF radiated energy between 5 and 200 MHz. This energy often peaks at the quarterwave resonant frequency of the person performing the test. Most human bodies self-resonate at between 35 and 45 MHz. Therefore, the peak radiated emissions from ESD also peak in this same frequency range.
The I/O cables are often resonant in this same range, and as a result receive significant amounts of the radiated ESD energy. Measured induced voltage levels across I/O cable termination loads show 600-V levels induced when the case is exposed to a 4- to 8-kV discharge. Such levels are well beyond the typical digital upset threshold value of 0.4 V. The typical induced pulse duration times are around 400 nanoseconds. Shielding the I/O cables and grounding at both ends with no internal signal lead exposure will reduce the levels 60 to 70 dB, yielding 0.3 V or less across the load.
EFTs also generate considerable radiated emissions that couple to cables and case circuitry. The coupling is similar to that experienced with ESD, but the discharge voltage levels are lower. The power supply EMC power-line filter provides the main protection to this transient. The CM line ground capacitors are normally the main devices used to shunt the transient currents to chassis and away from the internal circuitry. When this capacitance is limited by leakage current, the CM choke must provide more of the protection. This usually demands a special center tap common mode choke, where the center tap is shunted to chassis via a capacitor that complies with the leakage current limits. Typical inductance values are 1520 mH wound on a high-permeability ferrite core (typically 700010,000 µm).
Conclusion
The best approach is to treat all analog and digital circuits as high-frequency responsive and to use high-frequency design methods for cable shielding, PCB layout, and CM filtering methods. It is important to use full ground and power planes, even for analog circuits, to limit high-frequency common mode loops. Most transient threats are high frequency and cause significant radiated energy to be present. It is often this radiated energy that couples via I/O cables into a given piece of equipment. Often this energy comes in on analog leads, recouples to high-speed digital circuits, and causes them to respond adversely. By filtering all analog and digital leads, proper protection can be achieved.
Chris M. Kendall is principal consultant and CEO at CKC Laboratories, Inc. (Mariposa, CA).